Zero Error and Index Correction in Instruments

Physics · Units And Measurements · NEET

Zero error is the reading an instrument shows when it should read zero (jaws fully closed). You always SUBTRACT it: Correct reading = Observed reading - Zero error. Index correction is the same idea for an optical bench: it fixes the small gap between an object/pin's tip and the mark on the scale. Memory hook: "Zero error, subtract with sign" - a negative zero error subtracted becomes an ADD, so the diameter gets bigger.
Zero Error of a Screw Gauge (jaws closed)ReferencelinePositive zero errorzero is 3 divbelow line+3 x LC (reads high)Negative zero errorzero is 4 divabove line-4 x LC (reads low)Correct reading = Observed reading - Zero error (keep the sign)
When the jaws are closed, the position of the circular-scale zero against the reference line tells you the sign: zero below the line = positive error (reads high, subtract it); zero above the line = negative error (reads low, subtracting it adds back). Always use Correct = Observed - Zero error with the sign.

Your doubts, answered

Do I add or subtract the zero error?

Always SUBTRACT the zero error, keeping its sign: Correct reading = Observed reading - Zero error. If the zero error is negative, subtracting a negative number means you actually add its size. Example: if zero error = -0.004 cm, then Correct = Observed - (-0.004) = Observed + 0.004 cm. So the correct value comes out bigger.

What is the difference between positive and negative zero error?

Close the jaws with nothing in between. If the zero of the moving scale is AHEAD of the main-scale zero (reads a small positive value like +0.03 cm), that is a POSITIVE zero error - the instrument reads too high, so you subtract it. If the zero of the moving scale sits BEHIND the main-scale zero (below the reference line, appears to read a negative value), that is a NEGATIVE zero error - the instrument reads too low.

Is zero correction the same as zero error?

They are opposite in sign. Zero correction = -(Zero error). So Correct reading = Observed reading + Zero correction. Many NEET questions give you the zero error directly; just plug it into Correct = Observed - Zero error and the sign takes care of itself. Do not apply both a correction and a subtraction - use only one form.

What exactly is index correction?

On an optical bench you measure distances using the tips of pins or the position marks of lens/mirror holders. The tip of a pin usually does NOT sit exactly at the scale mark of its holder. Index correction is the fixed length you add or subtract to fix this gap so the measured object distance and image distance are true distances from the lens. It is zero error for the whole bench setup, not for one instrument.

How do I find the zero error of a screw gauge?

Close the studs with nothing between them and read the circular scale against the reference line. If the circular scale zero is above the reference line by 4 divisions and least count is 0.001 cm, the zero of the circular scale reads below the line, so zero error = -4 x 0.001 = -0.004 cm (negative). If it is below the line by n divisions, zero error = +n x least count (positive).

Which experiments need index correction and not just zero error?

Index correction is specifically used on the optical bench, for example when measuring the focal length of a lens or mirror using pins on an optical bench. A meter bridge uses end corrections (a similar idea, different name), a simple pendulum and a resonance tube do not need index correction. NEET 2026 asked exactly this: focal length of lenses using optical bench is the answer.

⚠️ The NEET trap
For zero error -0.004 cm, subtract 0.004 to get a smaller diameter.
Subtract WITH the sign: Correct = Observed - (-0.004) = Observed + 0.004, so the diameter increases.
🧠 A minus times a minus is a plus - negative zero error always makes the final reading LARGER, not smaller.

Real NEET questions

2018

A student measured the diameter of a small steel ball using a screw gauge of least count 0.001 cm. The main scale reading is 5 mm and zero of circular scale division coincides with 25 divisions above the reference level. If the screw gauge has a zero error of -0.004 cm, the correct diameter of the ball is:

A · 0.053 cm
B · 0.525 cm
C · 0.521 cm
D · 0.529 cm
Solution: Step 1: Main scale reading = 5 mm = 0.5 cm. Step 2: Circular scale contribution = 25 x least count = 25 x 0.001 = 0.025 cm. Step 3: Observed diameter = 0.5 + 0.025 = 0.525 cm. Step 4: Apply zero error with its sign: Correct = Observed - Zero error = 0.525 - (-0.004) = 0.525 + 0.004 = 0.529 cm. Answer: 0.529 cm (D). Note the trap - the negative zero error is ADDED, making the value bigger.
2025

The diameter of a spherical object is measured with a vernier caliper whose 10 vernier scale divisions (VSD) equal 9 main scale divisions (MSD). The least MSD is 0.1 cm and the zero of the vernier scale is at x = 0.1 cm when the jaws are closed. If the main scale reading for the diameter is M = 5 cm and the coinciding vernier division is 8, the measured diameter after zero-error correction is:

A · 4.98 cm
B · 5.00 cm
C · 5.18 cm
D · 5.08 cm
Solution: Step 1: Find least count. 10 VSD = 9 MSD, so 1 VSD = 0.9 MSD. LC = 1 MSD - 1 VSD = (1 - 0.9) x 0.1 = 0.01 cm. Step 2: Zero error = +0.1 cm (jaws closed, vernier zero sits at 0.1 cm, so it reads too high - positive). Step 3: Observed diameter = MSR + (VSD x LC) = 5 + (8 x 0.01) = 5 + 0.08 = 5.08 cm. Step 4: Correct = Observed - Zero error = 5.08 - 0.1 = 4.98 cm. Answer: 4.98 cm (A). Here a POSITIVE zero error is subtracted, so the value decreases.
2026

Which of the following measurements requires 'index correction'?

A · Measurement of resistance of a wire using meter bridge
B · Measurement of gravitational acceleration using simple pendulum
C · Measurement of focal length of lenses using optical bench
D · Measurement of speed of sound using resonance tube
Solution: Index correction fixes the small gap between the tip of a pin (or the optical centre of a lens/mirror) and the index mark on the optical-bench scale. It is applied when measuring focal length of lenses or mirrors on an optical bench. The meter bridge uses end corrections (a different name), and the pendulum and resonance tube do not need index correction. Answer: optical bench (C).

Solved Units And Measurements NEET PYQs

Try the real previous-year questions from this chapter — each with the answer and a full solution.

See all 30 Units And Measurements NEET PYQs ›
Next concept: What Is a Physical Quantity and Why We Measure ItKeep learning — 2 minFeeling ready? Solve the Units And Measurements NEET PYQs ›Or practice on your phone — get the free MedicNEET app ›

Frequently asked

Is zero error a systematic or random error?

Zero error is a systematic (instrumental) error. It shifts every reading by the same fixed amount and in the same direction, which is why it can be fully removed by simple correction. Random errors cannot be removed this way.

What is the formula for corrected reading?

Correct reading = Observed reading - Zero error, where the zero error is written with its own sign (positive or negative). Equivalently, Correct reading = Observed reading + Zero correction, where Zero correction = -(Zero error).

Can a good instrument have zero error?

Yes. Even a well-made vernier caliper or screw gauge can develop a small zero error from wear or manufacturing. That is why you always check the closed-jaw reading first and note the zero error before measuring.

Does zero error change the least count?

No. Zero error only shifts the reading up or down by a fixed amount. The least count depends only on the scale design (least count = 1 MSD - 1 VSD for vernier, pitch / number of circular divisions for screw gauge).

Why do we subtract a negative zero error?

Because the rule Correct = Observed - Zero error applies with sign. When zero error is negative, the instrument was reading too low, so subtracting a negative value adds it back, correctly increasing the reading.